Water Meter Test Bench Test Equipment single row DN15-50

Water Meter Test Bench Test Equipment single row DN15-50

This water meter test bench is used to test water meter according to relevant international standard ISO 4064, in order to find out if the water meter is Accepted or Rejected.

- optional 1 row, 2 rows, 4 rows
- 6 meters per row dn15, 5 meters per row dn20, 4 meters per row dn25 or dn32, 1 meter per row dn40 or dn50
- each row work independently
- test Qmax, Qn, Qt, Qmin or full error curve
- optional with/without pressure test
- packing: wooden case

test bench.pdf

 

Detailed information

- Pressure Test:
i) choose same size connectors, push control valve (1) clamp water meters;
ii) turn on ball valve (2), turn on ball valve (3), turn on ball valve (4), turn on ball valve (7);
iii) twist pressure test handle (6) to loose, wait 20 seconds, then turn off ball valve (7), turn off ball valve (4), then turn off ball valve (3);
iv) twist pressure test handle (6) to close slowly, watch pressure gauge (8), till the requested pressure, then turn off needle valve (5), wait the requested pressure test time;
v) finish pressure test, turn off ball valve (2), turn off ball valve (4), turn off ball valve (7), turn on ball valve (3), reduce pressure to zero;
 
if only do pressure test, then pull control valve (1), release water meters; or continue
 
- Flow Rate Error Test:
i) choose same size connectors, push control valve (1) clamp water meters;
ii) turn on ball valve (2), turn on ball valve (3), turn on ball valve (4), turn on ball valve (7);
iii) twist pressure test handle (6) to loose, wait 20 seconds, then turn off ball valve (7), turn off ball valve (4), then turn off ball valve (3);
iv) twist pressure test handle (6) to close slowly, watch pressure gauge (8), till the requested pressure, then turn off needle valve (5), wait the requested pressure test time;
v) finish pressure test, turn off ball valve (2), turn off ball valve (4), turn off ball valve (7), turn on ball valve (3), reduce pressure to zero;


WORKING PRINCIPLE:
        Permissible error:
        From Qt to Qmax both inclusive: ±2%
        From Qmin to Qt both exclusive: ±5%
       
Formula to calculate error:
        Error = (V2-V1)/V1 x 100% of which
        V2: Volume shown by the water meter(s)
        V1: Volume shown by the Volume tank

 

Please Attention:

1.If you want to get samples and catalogue from us, please contact us at info@sinoeco.com Normally, we need 7-15 days for making samples. and you need to pay for them. However, it would be highly appreciated if you can accept freight collect.
2.We normally need 20-35 days for production,however,please contact us if urgent
3.We ar OEM,ODM supplier,Wesupport FOB,CFR,DIF price term, and often ship from SHANGHAI or NINGBO sea port.
4.We encourage payment term:T/T or irrevocable L/C at sight, or 30% T/T deposit and 70% balance when receipt of advice of shipment.
5.Before installing the meter, we had better make sure that, it has been chosen correctly, check that the nominal diameter, flow rate, working temperature and pressure are compatible with the installation conditions.

Contact